LogTag Recorders LogTag Analyzer Temperature Recorder Manual de usuario Pagina 64

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 140
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 63
62 LogTag Analyzer
Automatically calculated
statistics
When a file is displayed either from downloading or from calling up a stored data file, the
software will automatically calculate and display some statistics about the recorded data.
The statistics that are calculated are shown on the corresponding Summary tab and are
also calculated for each sequence of data points between the inspection marks.
In This Section
Average Reading....................................................62
Standard Deviation.................................................63
Mean Kinetic Temperature......................................64
Degree Minutes......................................................65
Average Reading
The Average Reading is the arithmetic mean of the sequence of data points. It is
calculated by adding all the data point values together and then dividing the total by the
number of data points. Mathematically this formula can expressed as illustrated in the
following picture:
Where:
av the calculated arithmetic mean
n the number of data points to include in the calculation
t
i
a data point to include in the calculation
For example, if there were 5 data points to calculate the arithmetic mean of and they
were, 12.3, 15.9, 16.2, 14.7, 14.9, then the arithmetic mean would be 14.8.